Instrumentation for radioactivity: Difference between revisions
imported>Howard C. Berkowitz |
mNo edit summary |
||
(6 intermediate revisions by 2 users not shown) | |||
Line 1: | Line 1: | ||
{{ | {{PropDel}}<br><br>{{subpages}} | ||
{{ | |||
'''Instrumentation for radioactivity''' is of many types, due to different applications (e.g., analysis vs. safety), needs for portability, and the intensity and types of expected radiation. Instruments need to measure: | '''Instrumentation for radioactivity''' is of many types, due to different applications (e.g., analysis vs. safety), needs for portability, and the intensity and types of expected radiation. Instruments need to measure: | ||
<ref>{{cite web | <ref>{{cite web | ||
Line 10: | Line 9: | ||
}} </ref> | }} </ref> | ||
:* | :*Alpha particles | ||
:* | :*Beta particles | ||
:* | :*Neutrons | ||
:* | :*X-rays | ||
:* | :*Gamma rays | ||
Detectors are most likely to respond to the most energetic | Detectors are most likely to respond to the most energetic gamma rays, although some detectors are more selective, such as a zinc sulfide scintillation counter that responds only to alphas. A key, however, to measuring specific radiation types is shielding, such as subtracting, from the total radiation, that which is measured through a shield that only admits gammas, giving beta and alpha. | ||
Different types are needed, variously, for: | Different types are needed, variously, for: | ||
*Health risk to individuals | *Health risk to individuals | ||
*Analysis of nuclear materials | *Analysis of nuclear materials | ||
* | *Tritium survey | ||
*Mineral prospecting | *Mineral prospecting | ||
==Detector technology== | ==Detector technology== | ||
Line 29: | Line 28: | ||
| contribution = Module 1.13 Radiation Detector Theory Study Guide | | contribution = Module 1.13 Radiation Detector Theory Study Guide | ||
| title = Radiological Control Technician | | title = Radiological Control Technician | ||
| publisher = Office of Health, Safety and Security, | | publisher = Office of Health, Safety and Security, U.S. Department of Energy}}</ref> | ||
* | *Ionization detectors: Direct effects of radiation on the conductivity of electrically active material, such as Geiger-Mueller counters, portable ionization chambers, proportional counters | ||
* | *Excitation detectors: Indirect effects of ionizing radiation produce visible light that is measured: thermoluminescent dosimeters, scintillometers | ||
* | *chemical: photographic film badges | ||
*Other: activation foil, Cerenkov detector, | *Other: activation foil, Cerenkov detector, #bioassays|bioassays, #mass spetrometry|mass spectrometry | ||
===Ionization detectors=== | ===Ionization detectors=== | ||
In ionization detectors, the incident radiation creates ion pairs in the detector. Ionization detectors are common in measuring radioactivity, but | In ionization detectors, the incident radiation creates ion pairs in the detector. Ionization detectors are common in measuring radioactivity, but smoke detectors also make use of ionization. | ||
The substance to be ionized can be either gas (most common) or solid (semi-conductors). Gas filled | The substance to be ionized can be either gas (most common) or solid (semi-conductors). Gas filled | ||
chambers can be operated as either ion chambers, proportional counters, or Geiger- | chambers can be operated as either ion chambers, proportional counters, or Geiger- | ||
Mueller (GM) tubes. A typical solid ionization detector is a | Mueller (GM) tubes. A typical solid ionization detector is a germanium-lithium (GeLi) detector used in a | ||
multichannel analyzer<ref>DOE-HDBK-1122-99, p. 1.13-8</ref> Ionization detectors may be characterized as intended for personal, survey, or laboratory use, and if they use gaseous or solid detectors.<ref name=MR>{{citation | multichannel analyzer<ref>DOE-HDBK-1122-99, p. 1.13-8</ref> Ionization detectors may be characterized as intended for personal, survey, or laboratory use, and if they use gaseous or solid detectors.<ref name=MR>{{citation | ||
| url = http://www.iem-inc.com/prinsr.html | | url = http://www.iem-inc.com/prinsr.html | ||
Line 47: | Line 46: | ||
| publisher = Integrated Environmental Management, Inc.}}</ref> | | publisher = Integrated Environmental Management, Inc.}}</ref> | ||
==== | ====Personal dosimeter portable ionization chamber==== | ||
A gas-filled personal exposure instrument, the portable ionization detector is a small, air-filled container in which a quartz fiber is suspended, with a microscope that allows the shadow of the fiber to be read against a graduated scale. When the instrument is initialized, the fiber is charged to 200 volts, causing it to read a cumulative radiation exposure of zero. As the device is struck by ionizing radiation, the ions created in the air cause the fiber to discharge. | A gas-filled personal exposure instrument, the portable ionization detector is a small, air-filled container in which a quartz fiber is suspended, with a microscope that allows the shadow of the fiber to be read against a graduated scale. When the instrument is initialized, the fiber is charged to 200 volts, causing it to read a cumulative radiation exposure of zero. As the device is struck by ionizing radiation, the ions created in the air cause the fiber to discharge. | ||
Line 88: | Line 87: | ||
{{col-end}} | {{col-end}} | ||
====Smoke detector==== | ====Smoke detector==== | ||
Not all | Not all smoke detectors use ionization, but that is the most common method. In an ionization-based detector, an <sub>241</sub>americium source emits alpha particles that ionize air molecules in a chamber, giving a known current flow. The <sub>241</sub>Am is"embedded in a gold foil matrix...made by rolling gold and americium oxide ingots together to form a foil approximately one micrometer thick. This thin gold-americium foil is then sandwiched between a thicker (~0.25 millimeter) silver backing and a 2 micron thick palladium laminate. This is thick enough to completely retain the radioactive material, but thin enough to allow the alpha particles to pass." | ||
If smoke and combustion products enter the chamber, they reduce ionization and thus the conductivity of the air inside. The drop in current flow triggers the alarm. <ref>{{citation | If smoke and combustion products enter the chamber, they reduce ionization and thus the conductivity of the air inside. The drop in current flow triggers the alarm. <ref>{{citation | ||
| url = http://www.epa.gov/rpdweb00/sources/smoke_ion.html | | url = http://www.epa.gov/rpdweb00/sources/smoke_ion.html | ||
| publisher = | | publisher = Environmental Protection Agency | ||
| title = Ionization Technology}}</ref> | | title = Ionization Technology}}</ref> | ||
====Tritium survey==== | ====Tritium survey==== | ||
TRIUMF has built a tritium-in-air monitor that illustrates the problems of shielding and differential radiation measurement. "The measurement of tritium in the air presents a problem because the average energy of the beta particles emitted by tritium is so low (5.7 keV) that it is not possible to make a detector with walls thin enough to be penetrated by the tritium betas. Instead the air to be sampled is drawn through an ionisation chamber by a pump so that all of the beta particle energy is used to create ion pairs in the chamber. | |||
The instrument is compensated for use in gamma fields by having a second chamber in the monitor which is sealed and measures only the gamma field. The difference between the readings of the two chambers is a measure of the tritium concentration. "<ref>{{citation | The instrument is compensated for use in gamma fields by having a second chamber in the monitor which is sealed and measures only the gamma field. The difference between the readings of the two chambers is a measure of the tritium concentration. "<ref>{{citation | ||
| url = http://trshare.triumf.ca/~safety/EHS/rpt/rpt_6/node6.html | | url = http://trshare.triumf.ca/~safety/EHS/rpt/rpt_6/node6.html | ||
| publisher = | | publisher = TRIUMF | ||
| title = Tritium in Air Monitor}}</ref> | | title = Tritium in Air Monitor}}</ref> | ||
In the U.S. military, for applications such as surveys after | In the U.S. military, for applications such as surveys after nuclear accidents and for nuclear MASINT, while the T-446 and T-290A are tritium ''detectors'', the AN-PDR-74|/PDR-74 is a measurement instrument containing a "IM-246 light weight tritium air monitor to detect airborne radioactive gases. Also, the instrument is calibrated directly in terms of tritium activity but may also be used to detect other radiogases | ||
or to monitor gamma radiation if appropriate calibration factors are applied to the meter reading. The instrument | or to monitor gamma radiation if appropriate calibration factors are applied to the meter reading. The instrument | ||
is battery operated (D cells) and has an audible alarm when radioactivity exceeds a preset level." <ref>{{citation | is battery operated (D cells) and has an audible alarm when radioactivity exceeds a preset level." <ref>{{citation | ||
Line 111: | Line 110: | ||
| title = Nuclear Weapon Accident Response Procedures (NARP) | | title = Nuclear Weapon Accident Response Procedures (NARP) | ||
| author = Assistant to the Secretary of Defense (Atomic Energy) | | author = Assistant to the Secretary of Defense (Atomic Energy) | ||
| publisher = | | publisher = U.S. Department of Defense | ||
| url = http://www.fas.org/nuke/guide/usa/doctrine/dod/5100-52m/index.html | | url = http://www.fas.org/nuke/guide/usa/doctrine/dod/5100-52m/index.html | ||
}}, p. 5-A-1 to 5-A-2</ref> | }}, p. 5-A-1 to 5-A-2</ref> | ||
===Excitation=== | ===Excitation=== | ||
"In excitation detectors, the incident radiation excites the atoms of the detector material. | "In excitation detectors, the incident radiation excites the atoms of the detector material. | ||
The atoms give off the excess energy in the form of visible light."<ref>DOE-HDBK-1122-99, p. 1.13-8</ref> In some instruments, not usually considered radiation survey, the emitted light can be formed into images, as in | The atoms give off the excess energy in the form of visible light."<ref>DOE-HDBK-1122-99, p. 1.13-8</ref> In some instruments, not usually considered radiation survey, the emitted light can be formed into images, as in computed tomography. | ||
=====Thermoluminescent dosimeters===== | =====Thermoluminescent dosimeters===== | ||
Thermoluminescent dosimeter (TLD) badges, like film dosimeters, are issued to radiation workers and emergency responders who have the potential to be exposed to ionizing radiation. Thermoluminescent crystals inside the device are analogous to the film in film badges, except that radiation exposures cause them to glow, and the emitted light can be measured and exposure calculated from its intensity. TLDs are especially good for measuring low-level radiation exposure, in the 1-105 rem range. | Thermoluminescent dosimeter (TLD) badges, like film dosimeters, are issued to radiation workers and emergency responders who have the potential to be exposed to ionizing radiation. Thermoluminescent crystals inside the device are analogous to the film in film badges, except that radiation exposures cause them to glow, and the emitted light can be measured and exposure calculated from its intensity. TLDs are especially good for measuring low-level radiation exposure, in the 1-105 rem range. | ||
Line 133: | Line 132: | ||
{{col-end}} | {{col-end}} | ||
====Scintillometer==== | ====Scintillometer==== | ||
Like a TLD, the crystal in a scintillometer emits light when irradiated. Unlike a TLD, it emits only a burst when struck, and the intensity of the burst is low, requiring that the light be detected and recorded by electronics, such as a | Like a TLD, the crystal in a scintillometer emits light when irradiated. Unlike a TLD, it emits only a burst when struck, and the intensity of the burst is low, requiring that the light be detected and recorded by electronics, such as a photomultiplier tube. Scintillometers are most often used for detecting and measuring, but their "big brothers" in nuclear medicine, such as a single photon emission computed tomography (SPECT scanner), can create images. | ||
A basic field survey scintillometer, such as a military | A basic field survey scintillometer, such as a military PDR-77|AN/PDR-77, comes with a set of probes variously intended for alpha, beta/gamma, and low-energy X-ray radiation. The X-ray probe allows detection of plutonium and americium contamination. "Knowing the original assay and the age of the weapon, the ratio of plutonium to americium may | ||
be computed accurately and so the total plutonium contamination may be determined. <ref name=DoD3150.8-M>{{citation | be computed accurately and so the total plutonium contamination may be determined. <ref name=DoD3150.8-M>{{citation | ||
| title= DoD 3150.8-M, "Nuclear Weapon Accident Response Procedures (NARP)" | | title= DoD 3150.8-M, "Nuclear Weapon Accident Response Procedures (NARP)" | ||
| author = | | author = United States Department of Defense | ||
| url = http://www.dtic.mil/whs/directives/corres/pdf/315008m.pdf}}</ref> | | url = http://www.dtic.mil/whs/directives/corres/pdf/315008m.pdf}}</ref> | ||
Detector types in use include: | Detector types in use include: | ||
*''' | *'''Zinc sulfide (ZnS)''': detect alphas in the presence of other types of radiation by energy discrimination. A thin coating of zinc sulfide (a phosphor) is placed behind an alpha-transparent thin entrance window; alpha particles go through the window and produce measurable light flashes; higher-energy particles interact too quickly to produce the flash. A nonquantitative alpha detector using ZnS is called a '''spinthariscope''', and is used in introductory education. | ||
*''' | *'''Sodium iodide (NaI)''' for "low levels of gamma/x-ray radiation. These detectors typically read out in units of cpm, but with proper calibration and within proper energy limits, they may be used as a microroentgen ("micro-R") meter to measure low exposure rates. A sodium iodide detector is a solid chunk of material with an outer casing. The thickness of the casing prohibits detection of alpha and beta radiation". Large NaI scintillation crystals are the heart of the Anger camera that forms images in various nuclear medicine scanners. | ||
*'''Plastic scintillators''' are organic chemicals that, depending upon the material and way in which it is packaged, plastic scintillators can be used for alpha, beta, gamma or neutron detection. | *'''Plastic scintillators''' are organic chemicals that, depending upon the material and way in which it is packaged, plastic scintillators can be used for alpha, beta, gamma or neutron detection. | ||
===Chemical=== | ===Chemical=== | ||
Film dosimeters, or film badges, do not provide direct response, but are inexpensive and widely used for safety monitoring of personnel routinely exposed to radiation. The badge consists of several layers: | Film dosimeters, or film badges, do not provide direct response, but are inexpensive and widely used for safety monitoring of personnel routinely exposed to radiation. The badge consists of several layers: | ||
*Front, with identification information and a window for exposure | *Front, with identification information and a window for exposure | ||
Line 170: | Line 169: | ||
*Film badges may be left or lost at the site of the radiation accident | *Film badges may be left or lost at the site of the radiation accident | ||
{{col-end}} | {{col-end}} | ||
===Bioassays=== | ===Bioassays=== | ||
Bioassays, in the context of measuring human radioactive exposures, are analytical technologies that detect the type and amount of radionuclides in a urine sample to determine the amount of internal radionuclide contamination that a person has received during a radiological or nuclear incident. <ref name=CDC-GR-RW>{{citation | |||
| journal = Morbidity & Mortality Weekly Report | | journal = Morbidity & Mortality Weekly Report | ||
| title = CDC Grand Rounds: Radiological and Nuclear Preparedness | | title = CDC Grand Rounds: Radiological and Nuclear Preparedness | ||
Line 181: | Line 181: | ||
===Mass spectrometry=== | ===Mass spectrometry=== | ||
When used to measure human exposures, | When used to measure human exposures, mass spectrometry technologies detect the actual number of radionuclide atoms instead of the alpha, beta, or gamma emissions. <ref name=CDC-GR-RW/> | ||
==Field applications== | ==Field applications== | ||
===Health=== | ===Health=== | ||
According to the Federal Emergency Management Agency, there are three portable types of instruments to measure the dose received by individuals, each with advantages and disadvantages: portable ionization chambers, film badges, and thermoluminescent personal dosimeters. <ref name=FEMA-IS>{{citation | |||
| id = IS 301 | | id = IS 301 | ||
| date January 1998 | | date January 1998 | ||
Line 193: | Line 193: | ||
Portable detectors share the problems that they may become contaminated and artificially raise the exposure reading. Small units can be easily lost. | Portable detectors share the problems that they may become contaminated and artificially raise the exposure reading. Small units can be easily lost. | ||
Individual dose measurements for emergency responders is the responsibility of the Safety Officer in the | Individual dose measurements for emergency responders is the responsibility of the Safety Officer in the Incident Command System. | ||
===Detection of Nuclear Weapons and Materials=== | ===Detection of Nuclear Weapons and Materials=== | ||
Given the danger of smuggled nuclear weapons, as well as the | Given the danger of smuggled nuclear weapons, as well as the special nuclear materials (SNM) to build them, intense attention is being paid to detecting them. The problem divides roughly into systems practical for fixed or minimally mobile installation at transportation ports of entries, and for field searches for weapons development. <ref name=CRS>{{citation | ||
| title = Detection of Nuclear Weapons and Materials: Science, Technologies, Observations | | title = Detection of Nuclear Weapons and Materials: Science, Technologies, Observations | ||
| author = Jonathan Medalia | | author = Jonathan Medalia | ||
Line 209: | Line 209: | ||
#Nuclear resonance fluorescence imaging to identify materials based on the spectrum of gamma rays a nucleus emits when struck by photons of a specific energy. | #Nuclear resonance fluorescence imaging to identify materials based on the spectrum of gamma rays a nucleus emits when struck by photons of a specific energy. | ||
#The Photonuclear Inspection and Threat Assessment System to detect SNM up to 1 km away, unlike other systems that operate at very close range. It would beam high energy photons at distant targets to stimulate fission in SNM, producing characteristic signatures that may be detected. | #The Photonuclear Inspection and Threat Assessment System to detect SNM up to 1 km away, unlike other systems that operate at very close range. It would beam high energy photons at distant targets to stimulate fission in SNM, producing characteristic signatures that may be detected. | ||
==Laboratory analysis== | ==Laboratory analysis== | ||
There are limits to what can be determined with portable equipment. For more complex analysis, either a transportable laboratory needs to be brought to the site, or, if safety permits, to have representative samples taken to a laboratory. Analysis of radioactive trace elements, for example, can help identify the source of fuel for a given contamination incident. Some of the less portable,, but powerful instrumentation includes | There are limits to what can be determined with portable equipment. For more complex analysis, either a transportable laboratory needs to be brought to the site, or, if safety permits, to have representative samples taken to a laboratory. Analysis of radioactive trace elements, for example, can help identify the source of fuel for a given contamination incident. Some of the less portable,, but powerful instrumentation includes gamma spectrometry, low background alpha and beta counting and liquid scintillation counter for extremely low energy beta emitters such as tritium. | ||
The DoD directive makes the distinction clear that detection is harder than measurement, and the latter is necessary for MASINT: <blockquote>Nuclear radiation is not easy to detect. Radiation detection is always a multistep, highly indirect process. For example, in a scintillation detector, incident radiation excites a fluorescent material that de-excites by emitting photons of light. ... the quantitative relationship between the amount of radiation actually emitted and the reading on the meter is a complex function of many factors. Since those factors may only be controlled well within a laboratory. Such a laboratory either must be moved to the field, or samples brought to a fixed laboratory.<ref name=DoD3150.8-M/></blockquote> | The DoD directive makes the distinction clear that detection is harder than measurement, and the latter is necessary for MASINT: <blockquote>Nuclear radiation is not easy to detect. Radiation detection is always a multistep, highly indirect process. For example, in a scintillation detector, incident radiation excites a fluorescent material that de-excites by emitting photons of light. ... the quantitative relationship between the amount of radiation actually emitted and the reading on the meter is a complex function of many factors. Since those factors may only be controlled well within a laboratory. Such a laboratory either must be moved to the field, or samples brought to a fixed laboratory.<ref name=DoD3150.8-M/></blockquote> | ||
Line 217: | Line 218: | ||
==References== | ==References== | ||
{{reflist|2}} | {{reflist|2}}[[Category:Suggestion Bot Tag]] |
Latest revision as of 11:01, 1 September 2024
This article may be deleted soon. | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|
Instrumentation for radioactivity is of many types, due to different applications (e.g., analysis vs. safety), needs for portability, and the intensity and types of expected radiation. Instruments need to measure: [1]
Detectors are most likely to respond to the most energetic gamma rays, although some detectors are more selective, such as a zinc sulfide scintillation counter that responds only to alphas. A key, however, to measuring specific radiation types is shielding, such as subtracting, from the total radiation, that which is measured through a shield that only admits gammas, giving beta and alpha. Different types are needed, variously, for:
Detector technologyInstruments operate by one of four general principles: [2]
Ionization detectorsIn ionization detectors, the incident radiation creates ion pairs in the detector. Ionization detectors are common in measuring radioactivity, but smoke detectors also make use of ionization. The substance to be ionized can be either gas (most common) or solid (semi-conductors). Gas filled chambers can be operated as either ion chambers, proportional counters, or Geiger- Mueller (GM) tubes. A typical solid ionization detector is a germanium-lithium (GeLi) detector used in a multichannel analyzer[3] Ionization detectors may be characterized as intended for personal, survey, or laboratory use, and if they use gaseous or solid detectors.[4] Personal dosimeter portable ionization chamberA gas-filled personal exposure instrument, the portable ionization detector is a small, air-filled container in which a quartz fiber is suspended, with a microscope that allows the shadow of the fiber to be read against a graduated scale. When the instrument is initialized, the fiber is charged to 200 volts, causing it to read a cumulative radiation exposure of zero. As the device is struck by ionizing radiation, the ions created in the air cause the fiber to discharge. Some are direct, or self-reading, while others are indirect, or nonself-reading. There is also a variety of pocket ionization chambers that read at different rates (0.01-200 mR and 1-500 R). Pocket ionization chambers, primarily measure whole body gamma exposure (with some x-radiation).
Geiger-MuellerGeiger-Mueller counters are ionization detectors built around a gas-filled metal tube, with a wire charged to approximately +1000 volts down the center of the tube. When charged particles, such as ions or electrons from outside the tube, or X-rays or gamma rays that hit the outside of the tube and cause an electron to be emitted from the inside of the tube, the particles are attracted to the anode in the center. As particles traverse, the moving particles collide with more atoms in the gas, until the resulting "avalanche" produces sufficient electron flow to generate a measurable current pulse. Counter circuitry counts all pulses alike, with no differentiation by their energy or type other than they have enough energy to penetrate the tube. In practical counters, various shields or windows go around the tube to screen out less energetic particles; multiple tubes with different shielding can allow differentiation.
Smoke detectorNot all smoke detectors use ionization, but that is the most common method. In an ionization-based detector, an 241americium source emits alpha particles that ionize air molecules in a chamber, giving a known current flow. The 241Am is"embedded in a gold foil matrix...made by rolling gold and americium oxide ingots together to form a foil approximately one micrometer thick. This thin gold-americium foil is then sandwiched between a thicker (~0.25 millimeter) silver backing and a 2 micron thick palladium laminate. This is thick enough to completely retain the radioactive material, but thin enough to allow the alpha particles to pass." If smoke and combustion products enter the chamber, they reduce ionization and thus the conductivity of the air inside. The drop in current flow triggers the alarm. [5] Tritium surveyTRIUMF has built a tritium-in-air monitor that illustrates the problems of shielding and differential radiation measurement. "The measurement of tritium in the air presents a problem because the average energy of the beta particles emitted by tritium is so low (5.7 keV) that it is not possible to make a detector with walls thin enough to be penetrated by the tritium betas. Instead the air to be sampled is drawn through an ionisation chamber by a pump so that all of the beta particle energy is used to create ion pairs in the chamber. The instrument is compensated for use in gamma fields by having a second chamber in the monitor which is sealed and measures only the gamma field. The difference between the readings of the two chambers is a measure of the tritium concentration. "[6] In the U.S. military, for applications such as surveys after nuclear accidents and for nuclear MASINT, while the T-446 and T-290A are tritium detectors, the AN-PDR-74|/PDR-74 is a measurement instrument containing a "IM-246 light weight tritium air monitor to detect airborne radioactive gases. Also, the instrument is calibrated directly in terms of tritium activity but may also be used to detect other radiogases or to monitor gamma radiation if appropriate calibration factors are applied to the meter reading. The instrument is battery operated (D cells) and has an audible alarm when radioactivity exceeds a preset level." [7] Excitation"In excitation detectors, the incident radiation excites the atoms of the detector material. The atoms give off the excess energy in the form of visible light."[8] In some instruments, not usually considered radiation survey, the emitted light can be formed into images, as in computed tomography. Thermoluminescent dosimetersThermoluminescent dosimeter (TLD) badges, like film dosimeters, are issued to radiation workers and emergency responders who have the potential to be exposed to ionizing radiation. Thermoluminescent crystals inside the device are analogous to the film in film badges, except that radiation exposures cause them to glow, and the emitted light can be measured and exposure calculated from its intensity. TLDs are especially good for measuring low-level radiation exposure, in the 1-105 rem range.
ScintillometerLike a TLD, the crystal in a scintillometer emits light when irradiated. Unlike a TLD, it emits only a burst when struck, and the intensity of the burst is low, requiring that the light be detected and recorded by electronics, such as a photomultiplier tube. Scintillometers are most often used for detecting and measuring, but their "big brothers" in nuclear medicine, such as a single photon emission computed tomography (SPECT scanner), can create images. A basic field survey scintillometer, such as a military PDR-77|AN/PDR-77, comes with a set of probes variously intended for alpha, beta/gamma, and low-energy X-ray radiation. The X-ray probe allows detection of plutonium and americium contamination. "Knowing the original assay and the age of the weapon, the ratio of plutonium to americium may be computed accurately and so the total plutonium contamination may be determined. [9] Detector types in use include:
ChemicalFilm dosimeters, or film badges, do not provide direct response, but are inexpensive and widely used for safety monitoring of personnel routinely exposed to radiation. The badge consists of several layers:
After a timed period of exposure, the film is extracted and developed; the density of the developed film is proportional to the intensity of the radiation received. Some film-filter combinations may have different sensitivities in different parts of the film.
BioassaysBioassays, in the context of measuring human radioactive exposures, are analytical technologies that detect the type and amount of radionuclides in a urine sample to determine the amount of internal radionuclide contamination that a person has received during a radiological or nuclear incident. [10] Mass spectrometryWhen used to measure human exposures, mass spectrometry technologies detect the actual number of radionuclide atoms instead of the alpha, beta, or gamma emissions. [10] Field applicationsHealthAccording to the Federal Emergency Management Agency, there are three portable types of instruments to measure the dose received by individuals, each with advantages and disadvantages: portable ionization chambers, film badges, and thermoluminescent personal dosimeters. [11] Portable detectors share the problems that they may become contaminated and artificially raise the exposure reading. Small units can be easily lost. Individual dose measurements for emergency responders is the responsibility of the Safety Officer in the Incident Command System. Detection of Nuclear Weapons and MaterialsGiven the danger of smuggled nuclear weapons, as well as the special nuclear materials (SNM) to build them, intense attention is being paid to detecting them. The problem divides roughly into systems practical for fixed or minimally mobile installation at transportation ports of entries, and for field searches for weapons development. [12] A representative set of technologies includes:
Laboratory analysisThere are limits to what can be determined with portable equipment. For more complex analysis, either a transportable laboratory needs to be brought to the site, or, if safety permits, to have representative samples taken to a laboratory. Analysis of radioactive trace elements, for example, can help identify the source of fuel for a given contamination incident. Some of the less portable,, but powerful instrumentation includes gamma spectrometry, low background alpha and beta counting and liquid scintillation counter for extremely low energy beta emitters such as tritium. The DoD directive makes the distinction clear that detection is harder than measurement, and the latter is necessary for MASINT:
Detectors based on semiconductors, such as germanium, have better intrinsic energy resolution than scintillators, and are preferred where feasible for gamma-ray spectrometry. Neutron detection is improved by using hydrogen-rich scintillation counters, such those using a liquid rather than a crystal scintillation source. References
|